The trumpet-like configuration of the pneumotachometer head is designed to achieve laminar flow over a wide range of flows. On that account Lilly type pneumotachometers are mostly linear from 0 to 12 L/s.
The resistance to flow may increase due to accumulation of secretions or other contaminants, and from condensation of water vapor, changing their flow characteristics. The flow characteristics also vary with the viscosity of the gas measured, which varies with the gas mixture, and with gas temperature (ref. 1).
Ref. 1 - Measuring gas flow A synopsis of conditions which affect the measurement of gas flow with a pneumotachometer can be found in Appendix A, pp. 30-32 in: Quanjer PhH, Tammeling GJ, Cotes JE, Pedersen OF, Peslin R, Yernault JC. Lung volumes and forced ventilatory flows. Eur Respir J 1993; 6 suppl. 16: 5-40. Erratum Eur Respir J 1995; 8: 1629.